(Downloading may take up to 30 seconds.
If the slide opens in your browser, select File -> Save As to save it.)
Click on image to view larger version.

Fig. 2. Long-term time course of Ca2+-triggered granule fusion. Granule light scattering, a semi-quantitative measure of granule fusion, was measured at intervals after the addition of Ca2+ buffer to give the following final concentrations: 0.001 µM (IM buffer),
; 11.9 µM,
; 572 µM,
; 1151 µM,
. The decrease in light scatter characteristic of fusion in this system was complete at the two higher Ca2+ concentrations at the earliest interval (
3 minutes after Ca2+ buffer addition) measurable. No further change in light scattering was recorded over the remaining measurement interval. Additional Ca2+ buffer was added at the 65-minute time point, resulting in the following approximate final concentration levels: <0.001 µM (IM buffer),
; 572 µM,
; 1450 µM,
; 2680 µM,
. The granules were pelleted by centrifugation at the 110-minute time point, concentrating them in the well bottom. No change in absorbance, indicative of further fusion, was observed in either case. Error bars indicate s.d. (n=3).